L3/L3s/L6/L6s Split UV-Vis Beam Spectrophotometer
1. Split beam optical system
2. Automatic wavelength settings
3. 8 inch Color touch-screen LCD
4. Automatic 8 position cuvette holder
5. Automatic zero and full scale adjustment
6. Equipped with USB port
7. Optional thermal plotter
8. Optional software
Common features
1. The L series of spectrophotometers adopt split beam survey optical system, and blazed holographic gratings. They have outstanding test precision and very competitive prices.
2. 8-inch color touch screen LCD display, cutting-edge user interface, powerful functions, and easy operation.
3. The automated eight-position cuvettes holder facilitates the operation under each measurement mode.
4. With powerful functions, the equipment shows great performance in qualitative and quantitative testing, such as:
Full-spectrum scanning
Detailed spectrum scanning
Time-based kinetics determination
GOTO λ
Linear regression
Concentration direct reading
Peak/Valley detecting
Multi-wavelength measurement
5. The equipment is designed with sophisticated power protection system. With high capacity of internal memory, it can store testing results, scanned images, regression equations and correction data. Therefore, it follows a fast initialization when power on.
6. The instrument can be connected with dedicated printer, which can print testing results, or draw curve from spectral scanning, fixed wavelength time-based scanning, and linear regression.
7. There is a USB port on the device, and UVwin8 software is optional to order. The device can be connected to a PC, which can not only enhance the performance in data testing and spectrum scanning, but also expand the memory to save more testing results.
Specification
Models |
L3 |
L3S(Scanning) |
L6 |
L6S(Scanning) |
Optical system |
1200 lines/mm holographic grating, C-T monochromator, Split beam |
Display |
7 in. Touch Screen |
Wavelength Range |
325-1100 nm |
190-1100 nm |
Wavelength MPE |
±0.5nm |
Wavelength Repeatability |
≤0.2nm |
Transmittance MPE |
±0.5%T |
±0.3%T |
Transmittance Repeatability |
≤0.20%T |
≤0.15%T |
Transmittance Measuring Range |
0.0%-200.0% (T) |
Absorbance Measuring Range |
-0.301-4.000 (A) |
Concentration range |
0.000~9999(C) |
Spectrum Bandwidth |
2nm |
Stray Light |
≤0.05%T |
≤0.05%T |
≤0.03%T |
Drift |
≤0.0009A/0.5h
(500nm,2h warm up) |
≤0.0009A/0.5h
(250nm/500nm,2h warm up) |
Signal Noise |
100%T Noise≤0.20%T
0%T Noise≤0.10%T |
100%T Noise≤0.15%T
0%T Noise≤0.10%T |
Baseline Flatness |
|
±0.002A
(335-1090nm) |
|
±0.002A
(200-1090nm) |
Overall Dimensions |
550×430×200(mm) |
G.W. |
20kg |
23kg |
Software |
Optional |
Optional accessories
Thermal plotter
10mm cuvette holder
50mm cuvette holder
100mm cuvette holder
UV WIN8 software